The purpose of this study was to develop a fluorescence chromatography method for the detection of cartilage oligomeric matrix protein (COMP) in the auxiliary diagnosis of rheumatoid arthritis (RA). The principle of double antibody sandwich method was used to prepare immunochromatographic test strips, and the performance evaluation and methodological comparison were carried out. Through the detection of clinical samples, a receiver operating characteristic (ROC) curve was obtained, and the sensitivity, specificity, positive and negative predictive values of the test strip were calculated. The linear range was 0.39-50.00 ng/mL. The coefficients of variation inter and intra batches were less than 15%. The test strip was stable at 37℃ for 20 days, and the variation range of fluorescence signal intensity was within 15%. There was no cross reaction with rheumatoid factor (RF) and anti-cyclic citrulline peptide (anti-CCP) antibody. Forty-eight clinical serum samples were detected in parallel with ELISA kit, and the correlation was good. The test strip prepared in this study was used to detect the sample, the cut-off value of COMP between RA patients and healthy people was 22.55 ng/mL (sensitivity 0.821, specificity 0.842, positive predictive value 0.741, negative predictive value 0.895). At the same time, the same sample was tested with ELISA kit, the sensitivity and specificity of the two methods reached more than 80%. A quantitative COMP fluorescence chromatography test strip was developed, which has the advantages of celerity, simplicity and sensitivity, and may provide rapid auxiliary diagnosis for RA patients.
张永生,王云龙,王继创,李玉林,王晓军,王旭东. 软骨寡聚基质蛋白荧光免疫层析定量检测方法建立[J]. Chinese Journal of Biotechnology, 2022, 38(6): 2332-2341